Daily opening for clinical team - equipment, sterilization, and operatory prep.
Add it to Practice Flow and it becomes yours to edit, assign and schedule. It will start generating on its cadence, so the work comes back on its own.
Add this checklist Browse moreActivate main water supply to dental units.
Dental units cannot run without water, and finding that out with a patient seated costs the appointment. Turning it on first also lets pressure stabilize before anyone needs a handpiece.
1. Locate the master water valve, usually in the mechanical room or utility closet. 2. Open it fully. 3. Listen for the compressor to begin cycling. 4. Check one operatory for water flow before moving on.
Start central vacuum and compressor.
Suction and air are the two utilities that stop a procedure dead. Both need several minutes to reach working pressure, so starting them last is what makes the first appointment run late.
1. Switch on the central vacuum. 2. Switch on the air compressor. 3. Confirm the compressor reaches its cut-off pressure. 4. Drain any condensate from the compressor tank if your unit requires it.
Inspect autoclave water and power on.
A sterilizer that runs dry damages its heating element and takes the practice down to one unit, or none. Checking fluid takes ten seconds and prevents a repair bill and a cancelled afternoon.
1. Open the reservoir and check the level against the fill line. 2. Top up with distilled water only, never tap. 3. Power on and let it reach standby temperature while you do the rest of the room.
Power on radiographic equipment.
Sensors that fail at the moment of exposure mean a retake, which is a second dose to the patient and a delay to the schedule. Finding a dead sensor now means you can move that patient to a working room.
1. Power on each sensor and the imaging workstation. 2. Take a test exposure or run the software self-check. 3. Confirm the image appears in the patient management software. 4. Flag any unit that fails so the schedule can be moved.
Power on chairs, lights, handpieces.
Chairs, lights and handpieces each have a warm-up or self-test. Powering them together at the start surfaces a failure while there is still time to move a patient to another operatory.
1. Power on each chair and operatory light. 2. Run each handpiece briefly to confirm air and water. 3. Check the curing light. 4. Report anything that does not start.
Verify adequate gas levels.
An empty oxygen tank during a medical emergency is the worst possible time to discover it. Nitrous running out mid-procedure means an uncomfortable patient and a treatment you cannot finish.
1. Open both tank valves. 2. Read the gauges. 3. Replace or reorder any tank below the practice threshold, typically one quarter full. 4. Log the levels.
Fill with fresh solution and activate.
Ultrasonic solution loses effectiveness as it loads with debris. Reusing yesterday's solution means instruments look clean but are not, and that carries through the whole sterilization chain.
1. Drain the previous solution. 2. Refill with fresh solution at the labelled dilution. 3. Power on and run a degas cycle if your unit has one. 4. Record the change of solution.
Process first instrument load.
The first load sets the pace for the whole day. Starting it before patients arrive means instruments are ready when the schedule tightens, instead of the schedule waiting on the autoclave.
1. Load the pouched instruments without overpacking. 2. Select the correct cycle. 3. Start it and note the time. 4. Confirm the cycle completes and the indicators changed before unloading.
Purge standing water from all lines.
Water sits in the lines overnight and grows biofilm. Two minutes of flushing is the CDC recommended step that keeps what reaches a patient's mouth within safe limits, and it is a documented expectation in an inspection.
1. Run each line, including handpieces and air or water syringes, for two full minutes. 2. Run them into a sink or suction, not into the operatory. 3. Do every operatory, not just the ones in use today.
Stock gloves, masks, gowns, eyewear.
A missing glove size or a gown you have to go hunting for mid-procedure breaks aseptic technique or breaks the appointment. Stocking now is a two minute job that prevents both.
1. Check gloves in every size the team wears. 2. Check masks, eyewear and gowns. 3. Restock to par from the central supply. 4. Note anything running low for the order.
Lay out instruments and materials.
Setting up while the patient waits reads as disorganized and eats the appointment. Laying out against the actual procedure also catches a missing instrument while there is still time to sterilize one.
1. Read the first appointment for each operatory. 2. Lay out the tray for that specific procedure. 3. Confirm the materials, shades or implant components are on hand. 4. Escalate anything missing now.
Verify lab deliveries arrived.
A case that has not arrived is only a problem if you find out when the patient is in the chair. Checking two days ahead leaves room to call the lab and still make the appointment.
1. Open the lab case list for today and the next two days. 2. Confirm each case has physically arrived. 3. Call the lab on anything missing. 4. Tell the front desk about anything at risk so the patient can be called.
Check medications and AED status.
Emergency drugs expire and AED batteries die quietly. Both are checked in seconds and are the two items you cannot improvise during an actual emergency.
1. Confirm the AED status light shows ready. 2. Check the drug expiry dates in the kit. 3. Confirm the oxygen and the bag valve mask are present. 4. Replace anything expired the same day.
Check battery on handheld X-ray.
A portable unit that dies mid-day sends you back to a fixed room and disrupts whatever was scheduled there. Charging overnight or first thing avoids that entirely.
1. Check the battery indicator. 2. Place it on charge if it is below full. 3. Confirm it holds charge; a unit that will not hold is a service call, not a daily annoyance.
Check par levels and restock supplies.
Running out of pouches or indicators mid-day stops sterilization, which stops the schedule. Restocking at a calm moment is far cheaper than solving it under pressure.
1. Check pouches, wraps, chemical indicators and barriers against par. 2. Restock the sterilization area and each operatory. 3. Add anything below par to the supply order.